Moritani With Industry Since 1901 ISO14001 (Head Office, Osaka & Nagoya)

テクニカルデータ

techdata01.jpg

Frequency range: 1Hz to 40MHz
Bode 100 Technical Data (549.49 kB) 

Bode100の日本語での概要説明と簡単な使い方

GettingStarted.pdf
Signal Source Wave form: Sinusoidal signal
Frequency range 10 Hz to 40 MHz (standard frequency range)*
1 Hz to 40 MHz (extended frequency range)*
*frequency ranges software switchable

Signal Level -27dBm to 13dBm
0.01VRMS to 1VRMS (at 50 Ohm load)

Accuracy of the source level (23°C +/-5°C) +/-0.3 dB (1Hz to 1MHz)
+/-0.6 dB (1MHz to 40 MHz)

Frequency response of the source level 1Hz to 40MHz +/-0.3dB (referred to 100kHz)
Frequency accuracy (23°C +/-5°C) +/-15ppm (<1 year after calibration)
+/-25ppm (<3 years after calibration)
Source impedance: 50 Ohm
Return loss (1Hz to 40MHz) > 28dB (VSWR < 1.09)
Spurious signals < -55dBc (typical)
Harmonics < -55dBc (typical)
Connector: BNC
Inputs: CH1, CH2 Input Impedance (high): 1MOhm +/-2% (by design)
Input capacitance: 40 to 55pF
Input Impedance (low): 50 Ohm
Return loss for low input impedance (1Hz to 40MHz) > 25dB (VSWR < 1.12)
Noise floor for the gain measurement Conditions:
Resolution bandwidth = 10Hz, select internal reference, PSOURCE = 13dBm, 50O load at CH2, 20dB attenuators for CH1 and CH2
1Hz to 5kHz: -100dB (typical)
5kHz to 50kHz: -110dB (typical)
50kHz to 20MHz: -115dB (typical)
20MHz to 40MHz: -110dB (typical)
Connectors BNC
Receiver bandwidth: 1Hz to 3kHz
Input attenuator: 0 dB, 10 dB, 20 dB, 30 dB, 40 dB
Input sensitivity: 100 mV full scale (for input attenuator 0 dB)
Dynamic range: > 100 dB (at 10 Hz receiver bandwidth)
Gain error: < 0.1 dB (calibrated)
Phase error: < 0.5° (calibrated)
USB Interface Connector Type B
PC Requirements Minimum configuration: Pentium 500 MHz, 256 MB RAM, CD-ROM drive
Recommended configuration: Pentium 1 GHz, 256 MB RAM, CD-ROM drive
Interface: USB 1.1 or 2.0
Supported operating systems: Windows 2000
Windows XP
Windows XP 64 bit
Windows Vista
Windows Vista 64 bit
Power Requirements AC power adapter
Input voltage/frequency 100...240 V/47...63 Hz
DC power supply
Input voltage/output power +10...24 V/10 W
Supply current at 12V: 450mA (typical)
at 18V: 300mA (typical)
at 24V: 230mA (typical)
Low supply voltage shut-down 8.25V (typical)
Inrush current for a low impedance source at 12V 15A for 300µs (typical)
Inrush current for a low impedance source at 24V 30A for 300µs (typical)
Inner connector +10...24 V
Outer connector Ground
Inner diameter 2.5 mm
Outer diameter 5.0 mm
Environmental Requirements Temperature [Storage] -35...+60ºC/-31...+140ºF
[Operating] +5...+40ºC/+41...+104ºF
[For specifications] 23ºC ± 5ºC/73ºF ± 18ºF
Relative humidity [Storage] 20...90%, non-condensing
[Operating] 20...80%, non-condensing
General Dimensions (w × h × d): 26 x 5 x 26.5 cm
10.25 x 2 x 10.5 inch
Weight - Bode 100: < 2kg/4.4 lb
Weight - Accessories: < 0.5kg/1.1 lb
Absolute Maximum Ratings DC supply voltage +28V
DC supply reverse voltage (device doesn't work) -28V
Maximum AC input signals at CH1 or CH2 (high impedance) 50VRMS up to 1MHz
30VRMS 1MHz to 2MHz
15VRMS 2MHz to 5MHz
10VRMS 5MHz to 10MHz
7VRMS 10MHz to 40MHz
Maximum DC input signal at CH1 or CH2 (high impedance) 50V
Maximum input power at CH1 or CH2 (low impedance) 1W (=7VRMS)
Maximum possible return power at the source connector 0.5W